diff mbox series

[v5,2/4] iio: ABI: docs: Document Senseair Sunrise ABI

Message ID 20210909094537.218064-3-jacopo@jmondi.org (mailing list archive)
State Changes Requested
Headers show
Series iio: chemical: Add Senseair Sunrise CO2 sensor | expand

Commit Message

Jacopo Mondi Sept. 9, 2021, 9:45 a.m. UTC
Add documentation for the sysfs attributes of the sunrise_co2 driver.

Signed-off-by: Jacopo Mondi <jacopo@jmondi.org>
---
 .../sysfs-bus-iio-chemical-sunrise-co2        | 38 +++++++++++++++++++
 1 file changed, 38 insertions(+)
 create mode 100644 Documentation/ABI/testing/sysfs-bus-iio-chemical-sunrise-co2
diff mbox series

Patch

diff --git a/Documentation/ABI/testing/sysfs-bus-iio-chemical-sunrise-co2 b/Documentation/ABI/testing/sysfs-bus-iio-chemical-sunrise-co2
new file mode 100644
index 000000000000..ee7aeb11709b
--- /dev/null
+++ b/Documentation/ABI/testing/sysfs-bus-iio-chemical-sunrise-co2
@@ -0,0 +1,38 @@ 
+What:		/sys/bus/iio/devices/iio:deviceX/in_concentration_co2_calibration_factory
+Date:		August 2021
+KernelVersion:	5.16
+Contact:	Jacopo Mondi <jacopo@jmondi.org>
+Description:
+		Writing '1' triggers a 'Factory' calibration cycle.
+
+What:		/sys/bus/iio/devices/iio:deviceX/in_concentration_co2_calibration_background
+Date:		August 2021
+KernelVersion:	5.16
+Contact:	Jacopo Mondi <jacopo@jmondi.org>
+Description:
+		Writing '1' triggers a 'Background' calibration cycle.
+
+What:		/sys/bus/iio/devices/iio:deviceX/error_status_available
+Date:		August 2021
+KernelVersion:	5.16
+Contact:	Jacopo Mondi <jacopo@jmondi.org>
+Description:
+		Reading returns the list of possible chip error status.
+		Available options are:
+		- 'error_fatal': Analog front-end initialization error
+		- 'error_i2c': Read/write to non-existing register
+		- 'error_algorithm': Corrupted parameters
+		- 'error_calibration': Calibration has failed
+		- 'error_self_diagnostic': Internal interface failure
+		- 'error_out_of_range': Measured concentration out of scale
+		- 'error_memory': Error during memory operations
+		- 'error_no_measurement': Cleared at first measurement
+		- 'error_low_voltage': Sensor regulated voltage too low
+		- 'error_measurement_timeout': Unable to complete measurement
+
+What:		/sys/bus/iio/devices/iio:deviceX/error_status
+Date:		August 2021
+KernelVersion:	5.16
+Contact:	Jacopo Mondi <jacopo@jmondi.org>
+Description:
+		Reading returns the current chip error status.